Skip to main content
eScholarship
Open Access Publications from the University of California

UCLA

UCLA Previously Published Works bannerUCLA

Home use of a compact, 12‑lead ECG recording system for newborns

Abstract

BACKGROUND:An easy-to-operate ECG recorder should be useful for newborn screening for heart conditions, by health care workers - or parents. We developed a one-piece electrode strip and a compact, 12‑lead ECG recorder for newborns. METHOD:We enrolled 2582 newborns in a trial to assess abilities of parents to record a 12‑lead ECG on their infants (2-4 weeks-old). Newborns were randomized to recordings by parents (1290) or our staff (1292 controls). Educational backgrounds of parents varied, including 64% with no more than a high school diploma. RESULTS:For newborns randomized to parent recorded ECGs, 94% of parents completed a 10-minute recording. However, 42.6% asked for verbal help, and 12.7% needed physical help. ECG quality was the same for recordings by parents versus staff. CONCLUSIONS:By use of a one-piece electrode strip and a compact recorder, 87% of parents recorded diagnostic quality ECGs on their newborn infants, with minimal assistance.

Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.

Main Content
For improved accessibility of PDF content, download the file to your device.
Current View