Skip to main content
Download PDF
- Main
Comparison of SARS-CoV-2 Reverse Transcriptase Polymerase Chain Reaction and BinaxNOW Rapid Antigen Tests at a Community Site During an Omicron Surge
- Schrom, John;
- Marquez, Carina;
- Pilarowski, Genay;
- Wang, Chung-Yu;
- Mitchell, Anthea;
- Puccinelli, Robert;
- Black, Doug;
- Rojas, Susana;
- Ribeiro, Salustiano;
- Tulier-Laiwa, Valerie;
- Martinez, Jacqueline;
- Payan, Joselin;
- Rojas, Susy;
- Jones, Diane;
- Martinez, Daniel;
- Nakamura, Robert;
- Chamie, Gabriel;
- Jain, Vivek;
- Petersen, Maya;
- DeRisi, Joe;
- Havlir, Diane
- et al.
Published Web Location
https://doi.org/10.7326/m22-0202Abstract
Background
SARS-CoV-2 rapid antigen tests are an important public health tool.Objective
To evaluate field performance of the BinaxNOW rapid antigen test (Abbott) compared with reverse transcriptase polymerase chain reaction (RT-PCR) for detecting infection with the Omicron variant of SARS-CoV-2.Design
Cross-sectional surveillance study.Setting
Free, walk-up, outdoor, urban community testing and vaccine site led by Unidos en Salud, serving a predominantly Latinx community highly impacted by COVID-19.Participants
Persons seeking COVID-19 testing in January 2022.Measurements
Simultaneous BinaxNOW and RT-PCR from nasal, cheek, and throat swabs, including cycle threshold (Ct) measures; a lower Ct value is a surrogate for higher amounts of virus.Results
Among 731 persons tested with nasal swabs, there were 296 (40.5%) positive results on RT-PCR; 98.9% were the Omicron variant. BinaxNOW detected 95.2% (95% CI, 91% to 98%) of persons who tested positive on RT-PCR with a Ct value below 30, 82.1% (CI, 77% to 87%) of those who tested positive on RT-PCR with a Ct value below 35, and 65.2% (CI, 60% to 71%) of all who were positive on RT-PCR. Among 75 persons with simultaneous nasal and cheek swabs, BinaxNOW using a cheek swab failed to detect 91% (20 of 22) of specimens that were positive on BinaxNOW with a nasal swab. Among persons with simultaneous nasal and throat swabs who were positive on RT-PCR with a Ct value below 30, 42 of 49 (85.7%) were detected by nasal BinaxNOW, 23 of 49 (46.9%) by throat BinaxNOW, and 44 of 49 (89.8%) by either.Limitation
Participants were a cross-sectional sample from a community-based sentinel surveillance site, precluding study of viral or symptom dynamics.Conclusion
BinaxNOW detected persons with high SARS-CoV-2 levels during the Omicron surge, enabling rapid responses to positive test results. Cheek or throat swabs should not replace nasal swabs. As currently recommended, high-risk persons with an initial negative BinaxNOW result should have repeated testing.Primary funding source
University of California, San Francisco.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
For improved accessibility of PDF content, download the file to your device.
Enter the password to open this PDF file:
File name:
-
File size:
-
Title:
-
Author:
-
Subject:
-
Keywords:
-
Creation Date:
-
Modification Date:
-
Creator:
-
PDF Producer:
-
PDF Version:
-
Page Count:
-
Page Size:
-
Fast Web View:
-
Preparing document for printing…
0%