Single‐ and Bi‐component T2* analysis of tendon before and during tensile loading, using UTE sequences
Abstract
Background
To determine if the application of tensile force alters the single- or bi-component T2* values of human tendons as measured on a clinical MRI scanner with ultrashort echo time (UTE sequences and if single- or bi-component T2* values differ when measured with 2D-UTE, 3D-UTE, or 3D-UTE-Cones sequences.Methods
Ten tendons were imaged before and during the application of tension using various UTE sequences at 3 Tesla. Single and bi-component T2* analysis was performed pre- and posttension and compared with Bonferroni-corrected paired Wilcoxon tests.Results
Range of mean pre- and posttension T2* analysis values were: short T2* fraction (78.6-79.7% and 77.3-79.7%, respectively; P = 1.0 for all sequences), long T2* fraction (20.3-21.4% and 20.3-22.7%, respectively; P = 1.0 for all sequences), short T2* (0.9-1.0 ms and 0.9 ms, respectively; P = 1.0 for all sequences), long T2* (19.9-20.4 ms and 21.9-24.0 ms, respectively; P = 0.9 for 2D-UTE and P = 1.0 for 3D-UTE and 3D-UTE-Cones), and single-component T2* (2.3-2.5 ms and 2.5-3.2 ms, respectively; P = 1.0 for all sequences).Conclusion
No significant difference in single- or bi-component results was found after the application of tension to tendons. Results are similar regardless of UTE sequence used for acquisition.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
For improved accessibility of PDF content, download the file to your device.
Enter the password to open this PDF file:
File name:
-
File size:
-
Title:
-
Author:
-
Subject:
-
Keywords:
-
Creation Date:
-
Modification Date:
-
Creator:
-
PDF Producer:
-
PDF Version:
-
Page Count:
-
Page Size:
-
Fast Web View:
-
Preparing document for printing…
0%