Skip to main content
Download PDF
- Main
Simultaneous T1 and B1+ Mapping Using Reference Region Variable Flip Angle Imaging
Published Web Location
https://doi.org/10.1002/mrm.24904Abstract
Purpose
To present a new method that can simultaneously and efficiently measure T1 and B1 (+) maps using reference region variable flip angle (RR-VFA) imaging.Methods
Assuming T1 relaxation time in a reference region such as fat is well characterized, and the reference region sufficiently covers smoothly varying B1 (+) field inhomogeneity, B1 (+) maps can be measured from VFA images, conventionally used for T1 measurements. Fat-only images from two-point Dixon acquisitions were used to compute B1 (+) maps, and the B1 (+) maps were compared with ones using the double-angle method (DAM) in 22 breast MRI patients at 3T. Additionally, high spatial resolution VFA images were acquired to show T1 measurements with and without the RR-VFA B1 (+) correction in six patients.Results
RR-VFA is able to generate reliable B1 (+) maps, similar to those using the conventional DAM. This simultaneous T1 and B1 (+) mapping can also be used to reduce T1 estimation errors, where T1 maps have more uniform fibroglandular tissue T1 and better depiction of heterogeneous T1 of breast masses.Conclusion
A new method that can measure both T1 and B1 (+) maps based on Dixon VFA images is described, offering improved T1 quantification with no scan time penalty.Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.
Main Content
For improved accessibility of PDF content, download the file to your device.
Enter the password to open this PDF file:
File name:
-
File size:
-
Title:
-
Author:
-
Subject:
-
Keywords:
-
Creation Date:
-
Modification Date:
-
Creator:
-
PDF Producer:
-
PDF Version:
-
Page Count:
-
Page Size:
-
Fast Web View:
-
Preparing document for printing…
0%