Four-Dimensional Magnetic Resonance Imaging With 3-Dimensional Radial Sampling and Self-Gating–Based K-Space Sorting: Early Clinical Experience on Pancreatic Cancer Patients

Abstract

Purpose

To apply a novel self-gating k-space sorted 4-dimensional MRI (SG-KS-4D-MRI) method to overcome limitations due to anisotropic resolution and rebinning artifacts and to monitor pancreatic tumor motion.

Methods and materials

Ten patients were imaged using 4D-CT, cine 2-dimensional MRI (2D-MRI), and the SG-KS-4D-MRI, which is a spoiled gradient recalled echo sequence with 3-dimensional radial-sampling k-space projections and 1-dimensional projection-based self-gating. Tumor volumes were defined on all phases in both 4D-MRI and 4D-CT and then compared.

Results

An isotropic resolution of 1.56 mm was achieved in the SG-KS-4D-MRI images, which showed superior soft-tissue contrast to 4D-CT and appeared to be free of stitching artifacts. The tumor motion trajectory cross-correlations (mean ± SD) between SG-KS-4D-MRI and cine 2D-MRI in superior-inferior, anterior-posterior, and medial-lateral directions were 0.93 ± 0.03, 0.83 ± 0.10, and 0.74 ± 0.18, respectively. The tumor motion trajectories cross-correlations between SG-KS-4D-MRI and 4D-CT in superior-inferior, anterior-posterior, and medial-lateral directions were 0.91 ± 0.06, 0.72 ± 0.16, and 0.44 ± 0.24, respectively. The average standard deviation of gross tumor volume calculated from the 10 breathing phases was 0.81 cm(3) and 1.02 cm(3) for SG-KS-4D-MRI and 4D-CT, respectively (P=.012).

Conclusions

A novel SG-KS-4D-MRI acquisition method capable of reconstructing rebinning artifact-free, high-resolution 4D-MRI images was used to quantify pancreas tumor motion. The resultant pancreatic tumor motion trajectories agreed well with 2D-cine-MRI and 4D-CT. The pancreatic tumor volumes shown in the different phases for the SG-KS-4D-MRI were statistically significantly more consistent than those in the 4D-CT.