Aseptic Barriers Allow a Clean Contact for Contaminated Stethoscope Diaphragms

Abstract

Objective

To determine whether a single-use stethoscope diaphragm barrier surface remains aseptic when placed on pathogen-contaminated stethoscopes.

Methods

From May 31 to August 5, 2019, we tested 2 separate barriers using 3 different strains of 7 human pathogens, including extended-spectrum β-lactamase-producing Escherichia coli, methicillin-resistant Staphylococcus aureus, and vancomycin resistant Enterococcus faecium.

Results

For all diaphragms with either of the 2 barriers tested, no growth was recorded for any of the pathogens. Stethoscopes with aseptic barriers remained sterile for up to 24 hours. These single-use barriers also provided aseptic surfaces when stethoscope diaphragms were inoculated with human specimens, including saliva, stool, urine, and sputum.

Conclusion

Disposable aseptic diaphragm barriers may provide robust and efficient solutions to reduce transmission of pathogens via stethoscopes.