- Main
Compact in-vacuum gamma-ray spectrometer for high-repetition rate PW-class laser–matter interaction
Abstract
With the advent of high repetition rate laser facilities, novel diagnostic tools compatible with these advanced specifications are required. This paper presents the design of an active gamma-ray spectrometer intended for these high repetition rate experiments, with particular emphasis on functionality within a PW level laser-plasma interaction chamber's extreme conditions. The spectrometer uses stacked scintillators to accommodate a broad range of gamma-ray energies, demonstrating its adaptability for various experimental setups. In addition, it has been engineered to maintain compactness, electromagnetic pulse resistance, and ISO-5 cleanliness requirements while ensuring high sensitivity. The spectrometer has been tested in real conditions inside the PW-class level interaction chamber at the BELLA center, LBNL. The paper further details the calibration process, which utilizes a 60Co radioactive source, and describes the unfolding technique implemented through a stochastic minimization method.
Many UC-authored scholarly publications are freely available on this site because of the UC's open access policies. Let us know how this access is important for you.