Purpose
To examine the effects of volume and method on fluorescein tear breakup time (TBUT) values and to evaluate test efficacy in an independent sample free of selection bias.Methods
Subjects were assessed using a battery of dry eye tests (DETs). Efficacy study: Subjects were randomized to the DET, standard strip, and liquid NaFl on separate days. A masked examiner measured TBUTs from video recordings. Verification study: Subjects were investigated for efficacy using volumes of 5.0 and 2.0 μL mL of NaFl for TBUT.Results
Efficacy study: 46 subjects completed the study. Log-transformed TBUTs were significantly different, normal subjects versus dry subjects, for all 3 methods (all P values < 0.001). Area under the curves (AUCs), cut-points, sensitivity, and specificity were 1) DET: 0.873, 4.4 seconds, 0.97, and 0.67, respectively; 2) 2.0 mL: 0.901, 3.22 seconds, 0.90, and 0.87, respectively; and 3) standard strip: 0.912, 3.42 seconds, 0.97, and 0.80, respectively. Verification study: Data splitting analysis for the 2.0 μL data (n = 174 dry subjects and 97 normal subjects) generated an AUC of 0.917 and a cut-point of 6.05 seconds for a sensitivity of 0.87 and a specificity of 0.81. The 5.0 μL sample yielded an AUC of 0.940, with a sensitivity and specificity of 0.92 and 0.83, respectively, at a cut-point of 5.5 seconds.Conclusions
Little difference in TBUT was found using the 3 clinical methods with video recordings. Analysis using liquid NaFl suggests that the TBUT test has excellent diagnostic accuracy and that a cut-point of 5.3 to 6.0 seconds is the optimum to differentiate normals from persons with dry eye.