Purpose
To characterize the effect of hepatic vessel flow using 4-dimensional (4D) flow magnetic resonance (MR) imaging and correlate their effect on microwave ablation volumes in an in vivo non-cirrhotic porcine liver model.Materials and methods
Microwave ablation antennas were placed under ultrasound guidance in each liver lobe of swine (n = 3 in each animal) for a total of 9 ablations. Pre- and post-ablation 4D flow MR imaging was acquired to quantify flow changes in the hepatic vasculature. Flow measurements, along with encompassed vessel size and vessel-antenna spacing, were then correlated with final ablation volume from segmented MR images.Results
The linear regression model demonstrated that the preablation measurement of encompassed hepatic vein size (β = -0.80 ± 0.25, 95% confidence interval [CI] -1.15 to -0.22; P = .02) was significantly correlated to final ablation zone volume. The addition of hepatic vein flow rate found via 4D flow MRI (β = -0.83 ± 0.65, 95% CI -2.50 to 0.84; P = .26), and distance from antenna to hepatic vein (β = 0.26 ± 0.26, 95% CI -0.40 to 0.92; P = .36) improved the model accuracy but not significantly so (multivariate adjusted R2 = 0.70 vs univariate (vessel size) adjusted R2 = 0.63, P = .24).Conclusions
Hepatic vein size in an encompassed ablation zone was found to be significantly correlated with final ablation zone volume. Although the univariate 4D flow MR imaging-acquired measurements alone were not found to be statistically significant, its addition to hepatic vein size improved the accuracy of the ablation volume regression model. Pre-ablation 4D flow MR imaging of the liver may assist in prospectively optimizing thermal ablation treatment.