Purpose
The aim of this study was to describe the current state of science regarding independent external validation of artificial intelligence (AI) technologies for screening mammography.Methods
A systematic review was performed across five databases (Embase, PubMed, IEEE Explore, Engineer Village, and arXiv) through December 10, 2020. Studies that used screening examinations from real-world settings to externally validate AI algorithms for mammographic cancer detection were included. The main outcome was diagnostic accuracy, defined by area under the receiver operating characteristic curve (AUC). Performance was also compared between radiologists and either stand-alone AI or combined radiologist and AI interpretation. Study quality was assessed using the Quality Assessment of Diagnostic Accuracy Studies 2 tool.Results
After data extraction, 13 studies met the inclusion criteria (148,361 total patients). Most studies (77% [n = 10]) evaluated commercially available AI algorithms. Studies included retrospective reader studies (46% [n = 6]), retrospective simulation studies (38% [n = 5]), or both (15% [n = 2]). Across 5 studies comparing stand-alone AI with radiologists, 60% (n = 3) demonstrated improved accuracy with AI (AUC improvement range, 0.02-0.13). All 5 studies comparing combined radiologist and AI interpretation with radiologists alone demonstrated improved accuracy with AI (AUC improvement range, 0.028-0.115). Most studies had risk for bias or applicability concerns for patient selection (69% [n = 9]) and the reference standard (69% [n = 9]). Only two studies obtained ground-truth cancer outcomes through regional cancer registry linkage.Conclusions
To date, external validation efforts for AI screening mammographic technologies suggest small potential diagnostic accuracy improvements but have been retrospective in nature and suffer from risk for bias and applicability concerns.