Background
Powered hand exoskeletons are an emerging technology that have shown promise in assisting individuals with impaired hand function. A number of hand exoskeleton designs have been described in the literature; however, the majority have not been supported by patient-oriented criteria.Objective
The aim of this study was to define preliminary end-user needs and expectations for an assistive hand exoskeleton.Study design
Explorative interview and case series.Methods
Six clinicians and eight individuals with impaired hand function were interviewed in small groups or individually. A standardized list of questions was used to elicit feedback on specific design criteria or promote the discovery of new criteria. In addition, three participants with impaired hand function returned for a second session where hand characteristics, such as range of motion and force required to flex/extend fingers, were recorded to further quantify design requirements.Results
Interview responses indicated that there was general consensus among participants on criteria relating to important grasp patterns, grip strength, wear time, and acceptable bulk/weight. However, interview responses and hand characteristics also revealed important differences between individuals with impaired hand function.Conclusion
Qualitative and quantitative data were collected to develop an understanding of end-user design requirements for assistive hand exoskeletons. Although the data collected were helpful in identifying some preliminary criteria, differences between participants exist and identifying a universal set of criteria applicable across individuals with impaired hand function is challenging. This work reinforces the importance of involving users of rehabilitation technology in the device development process.